CAI 700 HCLD gas analyzer adopts the principle of chemical fluorescence method, which can quickly and accurately analyze NO/NOx and other gases. The instrument is designed based on 16-bit microprocessor, the input and output are 16-bit system, and the signal accuracy is high. It can be operated not only by instrument keyboard, but also by TCP/IP and RS-232C. The system can be set with 4-level password protection. The instrument has automatic calibration function and can also be calibrated remotely. How it works: CAI 700 HCLD Analyzer is based on chemical fluorescence principle and has two modes of NO and NOx. In NO mode, when NO reacts with O3 (ozone) in the gas sample to produce NO2, about 10% of NO2 is in an intensified state. When these excited molecules transition to the ground state according to formula (2), they emit light quantum hr with wavelength of 590 ~ 2500nm, and its intensity is proportional to the amount of NO. The NO concentration can be calculated by converting this light energy into electrical signal output by photocell. In NOx mode, the sample gas first enters the NOx conversion device. NOx in the sample gas includes NO and NO2, where NO2 is converted into NO. All NO is reacted and detected, and a DC current proportional to NOx is output. The digital panel table shows the concentration of NOx. The concentration of NO2 can be obtained by subtracting the concentration of NO from the concentration of NOx. NO + O 3 → NO 2 + O 2 (1) NO 2 → NO 2 + h v (2) Product features: Wide measuring range, 0 ~ 3-5000ppm The user can set the range of 4 gears Automatic calibration and automatic range switching Fast response speed Electronic control system for sample and ozone flow Automatic diagnosis of system state Remote monitoring and operation CE Mark and ETL Libed-complies with UL STD 61010-1, CAN/CSA C22.2 STD 61010.1 Main Applications: Continuous emission monitoring CEMS VOCs gas analysis Analysis of denitration efficiency Combustion efficiency detection Engine exhaust analysis... Technical indicators: Detector: Chemical fluorescence detector CLD Measuring range: 0 ~ 5000ppm NO/NOx users can set 4 ranges Response time: T90 < 2 seconds Repeatability: < 0.5% FS Zero/range drift: 1% FS/24 hours Flow control: EPC electronic flow control Sample flow rate: 1.5 ~ 2.5 LPM (optional 0.6 LPM) Ozone generator: ultraviolet lamp Air or oxygen requirements: 240cc/min, NOx < 0.01 ppm, 25psig, dew point < 10 ° C Output: TCP/IP, RS232, 4 ~ 20mA, 0-10V Alarm output: general fault, calibration error, high concentration alarm Condition diagnosis: Control voltage, temperature, pressure and flow Measurement cell temperature: HCLD 85 ° C, up to 100 ° C Ambient temperature: 5 C ~ 40 C Ambient humidity: < 90% RH Power supply: 115VAC/60Hz or 230VAC/50Hz, 200W Size: 5 Ω H x 19 W x 23 D (Inches) Weight: 45 lbs
|